22 
Balance Operating Menu 
 
 
The balance operating menu enables you to adapt the balance to your requirements. 
At the factory, we have set the codes for the balance operating program so that you normally will not need to 
make any changes. lf you have special operating conditions, change the codes in the balance operating 
program to meet your individual requirements. 
 
The 
factory settings
 in the balance operating program are marked with "*". Functions not marked by "*" can 
be set by entering the corresponding code (see page 26). 
lf you  need to change the factory settings,  make sure to  record the date. 
 
 
Weighing Mode: 
 
Adaptation to Ambient Conditions 
 
The balance is adapted to ambient conditions by changing the time allowed for weighings (see the Product 
Data Sheet). 
 
 Code Changes 
Very stable environment (shortest measuring time)  1  1  1   * 
Stable environment (short measuring time)  1  1  2   * 
Unstable environment (long measuring time)  1  1  3   
Very unstable environment (longest measuring time)  1  2  4   
 
 
Stability Range 
 
The stability symbol remains displayed for a weight 
deviation ± 
Code Changes 
0,25  digit  1  3  1   
0,5    digit  1  3  2   
1       digit  1  3  3   * 
2       digits  1  3  4   * 
4       digits  1  3  5   
8       digits  1  3  6   
 
 
Tare Parameters
 
 
The tare function is performed: 
 
 Code Changes 
Without stability control  1  5  1   
At stability  1  5  2   * 
 
 
Auto Zero Function 
 
When this function is activated, any zero point is eliminated so that a true zero indication is maintained. 
 
 Code Changes 
Auto-Zero on  1  6  1   * 
Auto-Zero off  1  6  2   
 
 
 
 
*  = factory setting, depends on balance model in some cases