28 Entris
®
BCE Operating Instructions
Operation
Requirements
The automatic or manual start of the isoCal function is set
in the menu (see Chapter “7.2.1 Switching the isoCAL
Function On or Off (Only Model i-1x)”, page 17).
The device is not located in the menu settings.
The load on the weighing pan remains unchanged for
2 minutes.
The load on the scales amounts to no more than 2% of the
maximum load.
The device does not register an input for 2 minutes.
Procedure
t If the automatic start of the isoCAL function is triggered:
y The [isoCAL] button flashes in the operating display.
y The operating display shows “CAL.” for 19 seconds.
y If no load change or no operation takes place on the
device before the expiration of the time display:
The isoCAL function starts.
t If the manual start of the isoCAL function is triggered:
y The [isoCAL] button flashes in the operating display.
t Press the [isoCAL] button.
y The isoCAL function starts.
8.6 Calibrating and Adjusting the Device
Internally (Only Model i-1x)
Requirements
The weighing pan is unloaded.
The operating display shows a stable weight value.
The internal calibration and adjustment function is set
(see Chapter “7.2.2 Setting Internal Calibration and
Adjustment (Only Model i-1x)”, page 17).
Procedure
t Press the [Zero] button.
t Press the [Adjust] button.
y The weight value is displayed.
t If the calibration function with subsequent automatic
adjustment is selected:
y The “CAL.RUN.” display appears in the operating display
during the process.
y The “CAL.END” display indicates the end of calibration.
t If the calibration function without subsequent automatic
adjustment is selected:
t Press the [Confirm] button.
y The calibration process starts.
y The “CAL.RUN.” display appears in the operating display
during the process.
y The calibration error is displayed.
t Press the [Confirm] button.
y The adjustment is performed.
y The “CAL.END” display indicates the end of calibration.
8.7 Externally Calibrating and Adjusting the
Device
Material Calibration and adjustment weight
Requirements
The weighing pan is unloaded.
The operating display shows a stable weight value.
The external calibration and adjustment function is set
(see Chapter 7.2.3, page 17).
The preset weight value for the adjustment weight or the
user-defined weight value for the adjustment weight is
selected (see Chapter 7.2.3, page 17).
Procedure
t Press the [Zero] button.
t Press the [Adjust] button.
t Place the indicated calibration and adjustment weight on
the balance.
y The weight value is displayed.
t If the calibration with subsequent automatic adjustment
function is selected and the calibration weight placed on
the device is within the specified limits:
y The adjustment process starts.
y If the “+” prefix was displayed: The applied weight is
too big.
y If the “-” prefix was displayed: The applied weight is
too small.
y The “CAL.END” message indicates the end of calibration.
t Remove the calibration weight.
t If the calibration without subsequent automatic adjustment
function is selected and the calibration weight placed on
the device is within the specified limits:
t Press the [Confirm] button.
y The adjustment process starts.
y If the “+” prefix was displayed: The applied weight is
too big.
y If the “-” prefix was displayed: The applied weight is
too small.
y The “CAL.END” message indicates the end of calibration.
t Remove the calibration weight.
8.8 Printing Results of the Calibration and
Adjustment Process
The results of the calibration and adjustment process can be
printed in a GLP printout.
Procedure
t Set the GLP printout in the menu (setting parameters,
see Chapter 7.3.7, page 22).
t Calibrate the device.
y Once the calibration is complete: The printing process starts.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius BCE1202-1 Series and is the answer not in the manual?

Sartorius BCE1202-1 Series Specifications

General IconGeneral
BrandSartorius
ModelBCE1202-1 Series
CategoryScales
LanguageEnglish

Related product manuals