36 Entris
®
BCE Operating Instructions
Technical Data
15.3.2 Power Supply Unit
Unit Value
Type: Sartorius power supply unit YEPS01-15V0W
Primary
Voltage V
AC
100 – 240 (±10 %)
Frequency Hz 50 – 60
Current consumption, maximum A 0.2
Secondary
Voltage V
DC
15 (±5 %)
Current, maximum A 0.53
Short-circuit protection Electronic
Protection class according to IEC 60950-1 II
Pollution level according to IEC 61010-1 2
Overvoltage category according to IEC 60664-1 II
Other data: See label on the power supply unit
15.4 Electromagnetic Compatibility
Interference resistance:
Suitable for use in industrial areas
Transient emissions:
Class B
Suitable for use in residential areas and areas that are connected to a low voltage network that also supplies residential
buildings.
15.5 Backup Battery
Unit Value
Lithium battery, type CR2032
Service life at room temperature, minimum Years 10
15.6 Materials
Housing Polybutylene terephthalate (PBT)
Control unit Glass
Draft shield Glass/polybutylene terephthalate (PBT)