32 Entris
®
BCE Operating Instructions
Malfunctions
10.2 Troubleshooting
Fault Cause Remedy Chapter, page
The operating display is
blank.
The device is disconnected. Check the connection to the power
supply.
The power supply unit is not
connected.
Connect the power supply cable to the
power supply.
The displayed weight value
changes constantly.
The installation site is unstable. Adjust the parameter for the ambient
temperatures.
Change the installation site.
A foreign object is positioned
between the weighing pan and
the housing.
Remove the foreign object.
The weight readout
displayed by the device
is obviously wrong.
The device has not been calibrated
and adjusted.
Calibrate and adjust the device. 8.4, 27
The device was not tared before
weighing.
Tare the device.
For a conformity-assessed
device: The [Invalid weight
value] display appears.
The cause of this fault is displayed
after pressing the [Change] key. At
the same time, the “Toggle between
weight units” function is locked.
ISOCAL.E.: The device needs to be
calibrated and adjusted.
Calibrate and adjust the device. 8.4, 27
WARMU.xx.: The device is in the
warm-up phase and has not yet
reached its operating temperature.
xx = remaining time in minutes
Comply with the warm-up time after
switching the device on.
15.7, 37
VALUE.ERR.: The displayed value is
invalid.
Set the device to zero.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius BCE3202-1 Series and is the answer not in the manual?

Sartorius BCE3202-1 Series Specifications

General IconGeneral
BrandSartorius
ModelBCE3202-1 Series
CategoryScales
LanguageEnglish

Related product manuals