Technical Data
15 Technical Data
15.1 Ambient Conditions
Unit Value
Installation site: For indoor use only, max. height above sea level m 3000
Temperature
Environment (metrological data)* °C +10 – +30
Environment °C +5 – +40
Storage and shipping °C -10 – +60
Relative humidity**
At temperatures up to 31 °C, non-condensing % 15 – 80
Then linear decrease from max. 80% at 31°C to max. 50% at
40°C
No heat from heating systems or direct sunlight
No electromagnetic fields
* For conformity-assessed (verified) balances in accordance with EU requirements, refer to the information on the balance.
** For conformity-assessed (verified) balances in accordance with EU requirements, the legal regulations apply.
15.2 Contamination Type, Overvoltage Category (Device)
Unit Value
Pollution level according to IEC 61010-1 2
Overvoltage category according to IEC 60664-1 I
15.3 Power Supply
15.3.1 Power Supply Device
Unit Value
Input voltage V
DC
15 (±10 %)
Power consumption, max. W 4
Only by Sartorius power supply unit YEPS01-15V0W