Weighing W
The basic weighing function is always
accessible and can be used alone or in
combination with application programs,
such as Counting, Checkweighing,
Weighing in Percent, etc.
Features
– Press ( to zero the weighing platform
– Press ) to store the weight on the
platform as tare
– Use the numeric keys to enter a tare
weight (press ) to save)
– Use a bar code scanner to enter
tare weight
– Tare container weight automatically
– Delete tare values by entering 0
(press ) to save)
– Press k to toggle between:
– Gross and net values, or
– 1st and 2nd weight unit, or
– Normal and 10-fold higher
resolution
Define the k key function in the
Setup menu, under:
Fn key function
– Weigh with two weighing platforms
– Individual ID codes for weight values
– Print weight values:
– Manually, by pressing p
– Automatically
(see “Data Output Functions”)
– With GMP-compliant format
(see “Data Output Functions”)
– Restore factory settings by selecting
the menu setting:
Application: Weigh:
Fty settings
Soft Key Functions
ID Enter up to four ID codes
for identifying results on
the printout
1st ID Save the value entered as
the first ID code
Preparation
§ Activate the Setup program:
M key
§ Select Application Parameters:
Press the
O soft key
§ Select “Weighing Only:”
Press the
O soft key
Weigh only
Minimum load for autotaring
1 digit
2 digits
5 digits
o 10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Autotare first weight
o Off
On
Factory settings
Yes
o No
o = factory setting
§ Save settings and exit Setup:
Press the M key or the
oo soft key
Automatic Taring
The first weight on the scale that
exceeds the preset minimum load is
stored in the tare memory at stability.
The values for subsequent loads are
stored as weight values. The scale
returns to the initial state when the
load is less than 50% of the minimum
load.
Configure in Setup under:
Application: Weigh:
Autotare 1st weight
Minimum Load
To tare container weights automatically,
you need to set a minimum load in the
Setup menu, under:
Application: Weigh:
MinL. autotar
The minimum load is defined in display
digits. There are ten levels to choose
from:
1 digit (no minimum load)
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
If the scale interval is 1 g, for example,
and the minimum load is set to 1000
digits (=1000 scale intervals), a load
of at least 1000 g is required for
autotaring.
Automatic Printing
The first weight value that exceeds the
minimum load is printed. Configure in
Setup under:
Device: Printout:
Once at stability
Weighing with Two Weighing
Platforms
Press the n key to toggle the display
between weighing platforms. Specify
one of the two platforms as the
“main scale” under
Device: Operat.:
Main scale
The display shows the readout from
the main scale when you switch on the
Combics. Press n to .toggle the
readout between platforms.
28
Operation