Service Menu
When the Service mode is activated, the service technician
can access additional menu items that are not visible when
the Service mode is deactivated.
The following advanced functions are available:
In the highest menu level, the last item is now:
Factory settings: all parametersFactory settings: all parameters
Factory settings: all parametersFactory settings: all parameters
Factory settings: all parameters
In the
Device parametersDevice parameters
Device parametersDevice parameters
Device parameters menu, the following items are
added to the menu, below
Code:Code:
Code:Code:
Code:
ServiceService
ServiceService
Service (service date)
MemoryMemory
MemoryMemory
Memory number (for a connected Alibi memory)
Terminal dataTerminal data
Terminal dataTerminal data
Terminal data (serial number and model of the
indicator)
The first item of the Device Parameters menu for „WP1 -
Internal“ is now:
ADC configuration.
The
Calibration/ adjustment Calibration/ adjustment
Calibration/ adjustment Calibration/ adjustment
Calibration/ adjustment menu in the Device
Parameters menu for „WP1 - Internal“ now contains the
following additional submenus:
Adjust without weights
Geographical data.
The CAL key function and Activate ext. adj. submenus
of the „Calibration/Adjustment“ menu also contain additi-
onal menu items.
If the COM1 or COM2 interface is configured for a second
weighing platform, „WP2“, the additions to the
„Calibration/Adjustment“ menu listed above for „WP1“
apply to the „WP2“ menu as well. Under certain conditions,
the „ADC Configuration“ menu is also available for WP2 (if
a weighing platform suitable for this purpose is connected
and a suitable transfer protocol is used).
In general, the software detects the parameters of the
connected load cell(s).
If the parameters of a given load cell cannot be detected
or cannot be changed, the parameters in question
might not be displayed. For example, in the „Calibration/
Adjustment“ menu, the „External calibration“ item is not
displayed if the connected weighing cell is not configured
for use in legal metrology; in other words, if it does not
have a „Legal metrology“ data record.