19
Internal Calibration/Adjustment
The menu code setting 1 9 3
must be selected in the Setup menu.
Inside the balance housing is a motori-
zed calibration weight which is applied
and removed automatically for internal
calibration.
Activate calibration: Press q
> The built-in calibration weight
is applied automatically
> The balance is calibrated
> If “Calibrate, then auto adjust in one
operation" is selected in the Setup
menu, the balance is now adjusted
automatically.
> The internal calibration weight
is removed.
Calibration and Adjustment Sequence
In the Setup menu, you can configure
whether:
calibration is always followed automati-
cally by adjustment (1 10 1; factory
setting), or
you have the choice of ending the
sequence or starting adjustment after
calibration (1 10 2)
If no difference is determined between
nominal and actual weights, you can
end the calibration routine following
calibration.
Two keys are active at this point:
q = start adjustment
c = end the sequence