Troubleshooting Guide
Error codes are shown on the main display for 2 seconds. The program then returns
automatically to the previous mode (e.g., weighing).
Display Cause Solution
No segments appear No AC power is available Check the AC power supply
on the display
The power supply is not Plug in the power supply
plugged in
h The load exceeds the balance Unload the balance
capacity
l or Err 54 Something is touching Move the object that is
the weighing pan touching the weighing pan
err 01 Data output not compatible Change the configuration in
with output format the operating menu
err 02 Calibration parameter Calibrate only when zero
not met; e.g.: is displayed
– balance not zeroed – Press w to tare the balance
– load on weighing pan – Unload the balance
err 10 The w key is blocked Press c to clear the tare
when there is data in the memory and release
second tare memory (net- the tare key
total); only 1 tare function
can be used at a time
err 11 Tare memory not allowed Press w
err 22 Weight is too light or, Increase the weight on the
when using an appli- balance
cation program, there is no
sample on the balance
err 30 Interface port for printer Reset the menu
output is blocked factory settings, or
Contact your local Sartorius
Service Center