Display Cause Solution
err 235 Connecting cable not  Connect the cable correctly
on model LE26P,   connected correctly
LE225D Cable connected to junction  Connect the equipment 
box on a different balance  correctly
The weight readout Unstable ambient conditions  Set up the balance 
changes constantly (excessive vibration or draft) in another area
at the place of installation
A foreign object is caught Remove the foreign object
between weighing pan and
balance housing
The weight readout  The balance was not  Calibrate/adjust the balance
is obviously wrong calibrated/adjusted before 
weighing
The balance was not  Tare or zero the balance 
tared/zeroed defore weighing before weighing
If any other errors occur, contact your local Sartorius Service Center!
Contact information:
Please point your Internet browser to:  http://www.sartorius.com
60