Setting IDs, Time, Date
and Display Brightness
Purpose
Configuring measurement environment parameters
for ISO/GLP-compliant data records.
Setting date and time (for ISO/GLP-compliant
records only).
Adapting display to ambient lighting conditions.
Features
Enter up to 8 characters to identify a measurement
series. Permissible characters include the numbers 0
through 9 and the dash or minus sign (“–"). A dash
is output as a space on printouts. Leading zeroes are
not output.
Date and time at beginning and end of ISO/GLP-
printouts.
Display brightness
1
):
0 = off; levels of brightness: 1 through 9
Key functions during configuration:
Activate input of IDs, time and date:
Switch the balance off and then on again by
pressing e; while all segments are displayed,
press the v key briefly
Scroll upward : Press q
Scroll to the right : Press r
Confirm input and toggle between IDs, time and
date: Press w
Save settings and exit menu:
Press and hold w (> 2 sec.)
1
) No display backlighting on models LE26P(-0CE)
and LE225D(-0CE)
30