Operating the Balance
Practical Example
Counting with: Preset Reference Sample Quantity Weighed In
(Parameter settings: Counting program: 2 1 4; Print all parameters: 7 1 2)
Step Key (or instruction) Display/Output
1. Select the counting application see “Preparation”
in the Setup menu
2. Prepare a container for the parts Place the empty container
on the balance
3. Tare the balance t
4. Place reference sample quantity Place the displayed number
on the balance (example: 10 pcs, of parts in the container
each weighing 2.148 g)
5. Initialize the balance start soft key v
(the number of digits following
the decimal point depends
on the balance model)
6. If necessary, increase Place parts in container
number of parts and update updat soft key v
the reference sample
(here: 7 more pieces)
7. Weigh uncounted parts Place parts to be counted
in container
8. If desired, print total piece count p
(here: 153 pcs)
wref=214800
.
wref=214300
.
wref=214300
.
nRef + 10 pcs
wRef + 2.14800 g
Qnt + 153 pcs
nref=10 pcs start
nref=10 pcs start
nref=10 pcs start