36
Weighing in Percent L
Purpose
This application program allows
you to obtain weight readouts
in percent which are in proportion
to a reference weight.
Available Features
Reference percentage “pRef”
loaded from long-term memory
when you turn on the balance
Optional balance configuration in
the Setup menu for automatically
initializing this application and
loading the most recent reference
percentage “pRef” entered with
reference weight “w100%” when
you turn on the balance
Minimum load checked
when the balance is initialized
(factory setting: 10 display
increments; can only be changed
by the Sartorius FastFactory)
Reference percentage “nRef”
entered using the numeric keys
Reference weight “Wxx%”
entered using the numeric keys
Storage parameter (rounding-off
factor) for storing the reference
weight “W100%” to calculate the
percentage can be configured
Configuration of decimal places
displayed with a percentage
Optional configuration for having
the reference weight “Wxx%”
and reference percentage
automatically output via the data
interface port after initialization of
the weighing-in-percent program
(print application parameters)
Long-term storage of
the last reference percentage
“pRef” entered
Long-term storage of the
last reference weight “W100%”
entered, by configuring automatic
initialization (auto-start)
Toggle between two weight units
by pressing w
Factory Settings
Auto-start application (automatic
initialization with reference
percentage and reference weight
loaded from long-term memory):
off (3 10 2)
Counting/percent parameter
(accuracy when storing reference
weights): display accuracy (3 5 2)
Number of decimal places displayed
in “weighing in percent” mode:
2 decimal places (3 6 3)
Printout application parameters
(automatic output of application
parameters): off 7 1 1)