Configuring the Balance
Entering User Data (Input)
Purpose
To display, input or change user
data.  You can block access to these
data by assigning a password.
Features
You can display, input or change
the following user data:
– Workstation number* for the
balance:  ID (balance ID; max.
20 characters)**
– Weighing series number, to
designate a series or lot:
L ID (lot ID; max. 20 characters)**
– Weight set number for calibration/
adjustment:  W ID (weight ID;
max. 14 characters)**
– Exact weight value for calibra-
tion/adjustment of the balance
(see the section on “Calibration/
Adjustment,” starting on page 24
in the chapter entitled “Operating
the Balance; in particular refer
to page 27)
– Password for access to the Setup
menu:  Input and Setup:
Menu (max. 8 characters)**
*** Only in conjunction with
ISO/GLP-compliant printouts
(see the section on “Setting
Parameters” in the chapter
entitled “Configuring the
Balance;” under menu code
number 8 10 x).
** A decimal point is displayed
together with its preceding digit
or character; it does, however,
count as a separate character.
This also applies when you
enter S ID and NUM as well
as to data entered via the
interface.
*** To delete user password:
Enter a decimal point using the
. key and confirm
Factory Settings
Password:  No designation
If no password has been assigned,
anyone can access the “Setup:
Input” and “Setup:  Menu” functions
without entering a password.
If you assign a password and then
forget what the word is, you can use
the General Password (see Appendix)
to access these menus.
Preparation
Display existing user data
● Select the Setup program:
Press M
> The soft keys Info, Menu and
Input are displayed in the text
line
● Select the user data input function:
Press the Input softkey v
> The password prompt is displayed
k If access is blocked by
a password:  enter the password
using the alphanumeric input keys
● Display user data:
Press the enter  passw.
soft key v
> The last 8 digits of a workstation/
balance number (ID no.), if
any ID is assigned, are displayed
in the measured value line
Enter/Change Password
● Select the Setup menu:
Press M
> The soft keys Info, Menu and
Input are displayed in the text
line
● Select the user data input function:
Press the Input soft key v
If you have already assigned
a password:
> The password prompt is displayed
k Enter the password
k Press the enter  passw.
soft key v
● Write down the password here:
Password = .............................
If you no longer remember the
password assigned:
k Enter the General Password
(see Appendix)
k Press the enter  passw.
soft key v
> The last 8 digits of a workstation/
balance number (ID no.), if any ID
is assigned, are displayed in the
measured value line
● Select password setting:  Press y
> Password is displayed
in the text line
> If a password exists, it is now
displayed in the measured value
line
● New password:  Enter the letters/
numbers for the new password
(8 characters max.)**
The password “none” means that
no password is stored.***
● Confirm input:  Press X
● Exit the Setup menu:
Press M
> Restart the application