EasyManua.ls Logo

Sartorius LP Models - Operating the Balance: Counting; Counting Application Overview; Counting Features and Factory Settings

Sartorius LP Models
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
33
Operating the Balance
Counting Z
Purpose
With the Counting program you
can determine the number of
parts that each have approximately
equal weight.
Available Features
Reference sample quantity “nRef”
loaded from long-term memory
when you turn on the balance
Optional balance configuration
in the Setup menu for auto-
matically initializing this applica-
tion and loading the most
recent reference sample quantity
“nRef” and average piece weight
“wRef” when you turn on the
balance (automatic initialization;
menu code 3 10 1)
Minimum load checked
when the balance is initialized
(factory setting: 10 display
increments; can only be changed
by the Sartorius FastFactory)
Reference sample quantity “nRef”
entered using the keys
Average piece weight “wRef”
entered using the keys
Storage parameter (level of
accuracy with which the average
piece weight “wRef” is stored)
for piece count calculation can
be configured
Optional configuration for having
the piece count and average
piece weight output automatically
via the data interface port after
initialization or reference sample
updating while running the
Counting program (print appli-
cation parameters)
Long-term storage of the last
reference sample quantity “nRef”
entered
Long-term storage of the
last average piece weight “wRef”
entered with the corresponding
reference sample quantity “nRef,”
by configuring automatic
initialization
Toggling between two weight
units by pressing w
Counting program initialized
again after using the balance for
weighing (after initialization)
Factory Settings
Auto-start application (automatic
initialization with reference sample
quantity and average piece weight
loaded from long-term memory):
off (3 10 2)
Counting/percent parameter
(accuracy when storing average
piece weights): display accuracy
(3 5 2)
Printout application parameters
(automatic output of application
parameters): off (7 1 1)

Related product manuals