EasyManua.ls Logo

Sartorius LP Models - Weighing in Percent Application Setup; Preparation and Parameter Settings; Weighing-in-Percent Parameter Configuration

Sartorius LP Models
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
Operating the Balance
Preparation
To calculate a value in percent, the
reference percentage must be
known. This value can be entered
into the weighing-in-percent program
in one of three ways:
The last reference percentage
entered is loaded and displayed
when you turn on the balance.
Place the corresponding weight
on the balance and initialize the
weighing-in-percent program;
With automatic initialization
switched on, the balance goes
into the “weighing in percent”
mode when you turn it on and
loads the last reference percent-
age entered as well as the
corresponding reference weight;
Enter the reference weight
using the numeric keys and store
it (W100% soft key).
Turn on the balance:
Press e
> All display segments light up
briefly
Select the Weighing-in-Percent
application in the Setup menu:
Press M
Select menu:
Press the menu soft key (w)
Set parameter 2 1 5:
See “Configuring the Balance”
Exit the Setup menu:
Press M
Setting Parameters for the Weighing-
in-Percent Application
k Access the Setup menu:
Select Menu
Set parameters for:
Storage accuracy:
3 5 Count/pct. param.
Decimal places displayed:
3 6 decimals f. calc.
Automatic initialization:
3 10 auto-start app.
Automatic output of parameters
to interface port:
7 1 Print app. param.
See “Configuring the Balance”
Exit the Setup menu:
Press M
Additional Functions
In addition to functions for:
alphanumeric input,
taring
(not during alphanumeric input),
printing,
you can also access the following
functions from this application:
calibration
(not during alphanumeric input),
setup,
turning off the balance.
Calibration/Adjustment
Press q
> See “Calibration/Adjustment”
for further instructions
Setup (setting parameters)
Press M
> See “Configuring the Balance”
for further instructions
Turning Off the Balance
Press e
> The balance shuts off
> The display goes blank

Related product manuals