Operating the Balance
Activating Data Output
You can define the data output
parameter so that output is activated
either when a print command
is received or automatically and
synchronous with the balance
display or at defined intervals
(see application program
descriptions and auto print settings).
Data Output by Print Command
The print command can be
transmitted by pressing p or by
a software command (Esc P).
Automatic Data Output
In the “auto print” operating mode,
data are output to the interface
port without a print command. You
can choose to have data output
automatically at defined print
intervals with or without the stability
parameter. Whichever parameter
you select, the data will be output
as the readouts appear on the
balance display. The display update
frequency depends on both the
model of the balance and the current
operating state.
If you select the auto print setting,
data will be transmitted immediately
the moment you turn on the balance.