Step Key (or instruction) Display
The printout header is
--------------------
printed 23.08.2005 11:25 *)
Model MA35M-230N
*) Date and time included only Ser. no. 99992581
if a Sartorius printer model Ver. no. 00-33-01
YDP20-0CE is used ID
--------------------
Temp. 130
,
C
Start W/O STABI.
End 10.0 min
IniWt + 2.036 g
--------------------
Current moisture loss and elapsed
time are displayed (in this example,
0.36% moisture after 0.3 min)
Drying stops automatically
after 10 minutes
The footer of the printout
--------------------
is printed 10.0 + 10.90 %M
FinWt + 1.814 g
Name:
--------------------
You can print the result
as often as you wish
10.0 + 10.90 %M
by pressing r
Printout when function canceled:
B 5.7 + 0.03 %M
(“B” stands for “Break”)
7. Clear the display
During and after the analysis you
can change the mode for display
and printout of results at any time
by pressing the
and keys.