Range 1, Range 2, Range 3
(RANGE 1, RANGE 2, RANGE 3)
The range limits are entered for the
individual ranges. The accuracy changes
when these limits are exceeded.
The following applies when entering
limits:
Range 1 < Range 2 < Range 3
< Max. cap.
This means that the weighing range can
be divided into a maximum of 4 ranges.
The resolution changes at intervals of
1, 2, 5, 10, 20 etc., where the lowest
resolution is the smallest scale interval
entered. Set ranges that are not
required for use to zero.
Available weighing units (menu item
UNITS)
This menu item is used to select the
weighing units that have been cleared
for use in weighing. All units marked
with a circle (o) have been cleared for
use, multiple selection is possible.
If you need to use this indicator as a
legal measuring instrument (legal for
trade), be sure you have selected a
permissible unit.
Save parameters (menu item SAVE)
The ADC configuration data is saved
once at the end of defining the settings
using the SAVE function.
Testing and configuration
for operation in legal metrology
A metrology plate is included in the
scope of supply of the indicator.
Once ADC configuration is complete,
record the metrological data for all
ranges on the metrology plate. Attach
the plate underneath the display and
cover with the supplied waterproof foil.
Under menu item 1.7, check that only
authorized weight units can be selected.
10
Menu Structure for ADC Configuration
The setup menu for WP1 (ā€œWP-1ā€) can be extended to include the following additional setting options for ADC configuration:
ADC CON ADC configuration
STAND. Standard configuration
RANGES Range selection
SINGLE WEIGHTS Single range mode
D Scale interval
Enter value
MAX Max. cap.
Enter value
MULT.INT WEIGHTS Multi-interval scale
D Scale interval
Enter value
RANGE 1 Range limit 1
Enter value
RANGE 2 Range limit 2
Enter value
RANGE 3 Range limit 3
Enter value
MAX Max. cap.
Enter value
MULT.R. WEIGHTS Multiple-range scale
D Scale interval
Enter value
RANGE 1 Range limit 1
Enter value
RANGE 2 Range limit 2
Enter value
RANGE 3 Range limit 3
Enter value
MAX Max. cap.
Enter value