EasyManuals Logo

Sartorius MIS2 User Manual

Sartorius MIS2
107 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #91 background imageLoading...
Page #91 background image
91
Maximum readability 15,000 display steps (not in legal metrology)
Accuracy class l, m (in the design...-CE)
Number of scale values 3,000e or 2+ 3.000e according to EN 45501
Digital protective interface According to EN45501
Data interfaces 2, optional
Display 14-segment backlit display
Housing:
Material Stainless steel, AISI 304 (1.4301)
Protection class: According to EN60529 IP65
Ambient conditions:
Operating temperature range -10°C to +40°C (+14°F to 104°C)
Humidity Maximum relative humidity 80% for temperature up to 31°C (~88°F);
linear decrease down to 50% for relative humidity at 40°C (+104°F)
Protection class of the housing
in accordance with EN 60529 IP 65
Pollution degree 2 Normally only nonconductive pollution occurs. Temporary conductivity caused
by condensation is to be expected.
Ratings:
Power supply 100-240 VAC (-15/+10%), 50-60 Hz, 17 W/23 VA max.
Transient overvoltage Overvoltage category II acc. to IEC 60364-4-443
Operation using protective extra low voltage See instruction manual for Option L8 (24-volt module)
DC supply 22.8 ... 26.7 V (optional: 21.6 ... 26.7 V); 12 VA max.
AC supply 22.8 ... 26.7 V, 50-60 Hz, 12 VA max.
Operation with rechargeable battery See Sartorius Installation Instructions for Option L9
Operation via built-in or external rechargeable battery (only available as an option
that must be ordered with the scale)
Emissions Acc. to EN613-1 (IEC 61326-1)
Group 1, Class B, suitable for use in domestic establishments and establishments
directly connected to a low-voltage power-supply network that supplies buildings used
for domestic purposes
Immunity to interference: Acc. to EN61326-1):
Immunity test requirements for equipment intended for use in industrial locations (Table 2)
Electrical safety Acc. to EN 61010-1 (IEC 1010-1), EN 60950 (IEC 950)
Temperature range -10°C to +40°C.
Power supply 100 – 2 40 VAC (–15/+10 %), 50 – 60 Hz, max. 17 W/23 VA
DC Supply optional 24 VDC (± 5%), max. 12 VA
AC Supply optional 24 VAC (± 5%), 50–60 Hz, max. 12 VA
Battery operation Via internal battery (only available as an option directly with order)
Interference According to EN61326-1 Class B (IEC 61326-1)
Interference resistance According to EN61326-1, industrial areas (IEC61326-1)
Electrical safety According to EN61010-1 (IEC1010-1)
Overview

Other manuals for Sartorius MIS2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius MIS2 and is the answer not in the manual?

Sartorius MIS2 Specifications

General IconGeneral
BrandSartorius
ModelMIS2
CategoryAccessories
LanguageEnglish

Related product manuals