EasyManuals Logo

Sartorius MIS2 User Manual

Sartorius MIS2
107 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
Key ) - > 2 sec Function
Allocation
Purpose
The key ) - > 2 sec is usually used
for the calibration/adjustment function.
The following additional functions can
be allocated to the key when the service
mode is activated:
External linearization with default
weights (menu item 1.9.6)
External linearization with the lineariza-
tion weights (menu item 1.9.7) entered
under menu item 1.18
Setting the preload (menu item 1.9.8)
Clearing the preload (menu item 1.9.9)
!Once linearization has been completed,
or after a preload has been set or
cleared the function of the key )- > 2
sec must be reallocated back to its nor-
mal function in the Setup menu (e.g.
external calibration/adjustment with
default weights)
1717
Menu structure for key ) - > 2 sec function allocation
1. 9. Calibration, adjustment
1. 9. 1 External calibration/adjustment with default weights (service mode not required)
1. 9. 3 External calibration/adjustment with user-defined weights (entered under 1-18), (service mode not required)
1. 9. 6 External linearization with default weights
1. 9. 7 External linearization with user-defined weights (entered under 1-18)
1. 9. 8 Set preload
1. 9. 9 Clear preload
1. 9. 10 Key blocked

Other manuals for Sartorius MIS2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius MIS2 and is the answer not in the manual?

Sartorius MIS2 Specifications

General IconGeneral
BrandSartorius
ModelMIS2
CategoryAccessories
LanguageEnglish

Related product manuals