9
when mounting the tip as this can damage
the pipette.
- Avoid rough handling and do not drop the
pipette.
- Avoid exposing the unit to extreme
temperature changes, humidity and dust
(operating temperature from 15°C to 4C).
11. Maintenance
mLINE pipettes have been designed for easy
in-house service. If the pipette is used daily it
is recommended to clean/decontaminate and
check the performance of the pipette every
three months. Sartorius also provides complete
repair and recalibration service for your pipette
including a service report and performance
certificate.
NOTE: Please make sure that the pipette has
been decontaminated before you send it to us
or your local representative. Please advise if any
hazardous material has been used with your
pipette.
NOTE: The use of the tip cone filters may prolong
the service interval. Change the filters regularly.
NOTE: It is recommended to always use gloves
when cleaning the pipette.
11.1. Daily cleaning of the outer
surface of the pipette
Your mLINE pipette should be checked every day
for cleanliness. To clean and decontaminate the
outer surface of your pipette use ethanol (70%),
isopropanol (60%) or a mild detergent and a soft
lint-free cloth.
Gently clean the surface of the pipette with
moistened cloth and wipe dry. Pay special
attention to the tip cone. Change the tip cone
filter if needed (Chapter 8.3.).
11.2. Cleaning the lower
part of the pipette
If your pipette is in daily use it is recommended
to clean/decontaminate and grease the pipette
every three months. It is recommended to send