Cubis MSE Operating Instructions  25
Getting Started
Semi-microbalances: Attaching the Display and Control Unit to the Electronics 
Module (MSE225..., MSE125... models)
The display and control unit can also be attached to the electronics module if 
required for operation�
t  Turn over the balance and place it on a soft surface�
Remove the connection cable from the cable channel:
t  Remove two screws from beneath the weigh cell and detach the plate�
t  Remove the connection cable plug� 
t  Then reattach the plate to the slot� 
t  Remove the display and control unit from the weigh cell:  
Remove two retaining screws�
t  Remove the display and control unit�