Operation
  Cubis MSE Operating Instructions  31
Level the Balance (Q-Level)
It is essential for exact weighing results that the balance is absolutely level� The 
front leveling feet can be used to level out small tilts in the floor� An integrated 
sensor detects the alignment of the balance and triggers a warning message when 
leveling is required� When “ " flashes on the display, you must level the balance� 
Leveling with the inclination sensor:
  Step  Press key  Display 
   1� 
 flashes on the display:     flashes 
 
  Start leveling�  *  SCREW IN
 
    2�  Fully screw in both back leveling feet   
  (depends on the model) and confirm�  *  TURN
    3�  Directional arrow on the right of TURN:  
  Turn right leveling foot�     TURN
      Directional arrow on the left of TURN:  
  Turn left leveling foot�    TURN
 
Directional arrow 
:
 Turn the respective leveling  
foot to the right until the  
directional arrow disappears�
Directional arrow  :  Turn the respective leveling  
foot to the left until the  
directional arrow disappears�      SCREW OUT
    4�   Screw out both back leveling feet  
(depends on the model) until they  
touch the setup surface�      *  0.0g
Leveling the balance with motorized leveling feet (optional):
t   To start the leveling function, press the * key�       flashes 
AUT.LEV
y   You can continue when the balance displays a weight value�
3
  Each time the balance is leveled, the balance must be calibrated/adjusted 
(see Calibrating/Adjusting the Balance)�
Turning the ionizer on/off (optional)
Purpose: The electrostatically charged weighing sample is neutralized as much as 
possible in the air flow of the ionizer�  
If your balance is equipped with an ionizer module, you can use the * key to 
turn the ionizer on/off�
Preparation
t  Set the intensity of the ionization in the menu� 
 Menu setting: extras: device: ionisat.: *Off, Soft intensity, Normal 
intensity
 or Strong intensity