94  Cubis MSE Operating Instructions
Technical Specifications
Model-specific Data
Verified Models with EC Type Approval Certificate: Micro- and ultramicrobalances
Model   MSE6.6S-0CE MSE2.7S-0CE MSE3.6P-0CE
Accuracy class*    k k k
For verified models: EC Type Approval Certificate D09-09-015, Type: MSX
Scale interval d*  mg  0�001  0�0001  0�001/0�002/0�005
Weighing capacity max*  g  6�1  2�1 1�1/2�1/3�1
Calibration value e*  mg  1  1  1
Min� load min*  mg  0�1  0�01  0�1
Tare equalization range (subtractive)    < 100 % from max� weighing capacity
Application range according to DIR*  g  0�001 – 6�1  0�001 – 2�1  0�001 – 3�1
Optimal starting point  
of the measuring range**  mg  0�82  0�082  0�82
Typical stabilization time  s  < 5  < 7  < 5
Typical measurement time  s  < 8  < 10  < 8
External standard calibration value  
(min� accuracy class)  g  5  2  3
Application range (temperature)     With “isoCAL” function: +5…+40°C |  
Without “isoCAL” function: +15 … +25°C
Display result (depending on the set filter level)  By selection of 1 of 4 optimized filter levels
Weighing pan size d  mm 30  20 30
Weighing chamber height  
(draft shield DM)  mm  70  70  70
Protection     Protected against dust and water
* DIR  = Directive 2014/31/EU on non-automatic weighing instruments used within the European Economic Area 
**    =  According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing 
range� Typically the value can be higher depending on the setup location and the ambient conditions�