Technical Specifications
Model-specific Data
Verified models with EC Type Approval Certificate: Precision balances 
Models    MSE5203S-0CE  MSE5203P-0CE  MSE3203S-0CE MSE3203P-0CE
Accuracy class*    k  k  k k 
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d*  mg  1  1/2/5  1  1/10
Weighing capacity max*  g  5,200  1,200/2,400/5,200  3,200  1,010/3,200
Calibration value e*  mg  10  10  10  10
Min� load min*  mg  100  100  100  100
Tare equalization range (subtractive)    < 100% from max� weighing capacity
Application range according to DIR*  g  0�1 – 5,200  0�1 – 5,200  0�1 – 3,200  0�1 – 3,200
Optimal starting point  
of the measuring range**  g  0�82  0�82  0�82  0�82
Typical stabilization time  s  < 1  < 1  < 1  < 1
Typical measurement time  s  < 2  < 2  < 2  < 1�5
External standard calibration value  g  5,000 (E2 5,000 (E2 2,000 (E2  2,000 (E2) 
(min� accuracy class)
Application range (temperature)    With “isoCAL” function: +5 ��� +40 °C  
    Without “isoCAL” function: +15 ��� +25 °C
Display result  
(depending on the set filter level)    0�1 – 0�4
Weighing pan size (W + D)   mm  140 + 140
Weighing chamber height (draft shield DE)
 mm  172
Protection     Protected against dust and water
Models    MSE2203S-0CE  MSE2203P-0CE MSE1203S-0CE
Accuracy class*    k  k k
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d*  mg  1  1/10  1
Weighing capacity max*  g  2,200  1,010/2,200  1,200
Calibration value e*  mg  10  10  10
Min� load min*  mg  100  100  100
Tare equalization range (subtractive)    < 100% from max� weighing capacity
Application range according to DIR*  g  0�1 – 2,200  0�1 – 2,200  0�1 – 1,200
Optimal starting point  
of the measuring range**  g  0�82  0�82  0�82
Typical stabilization time  s  < 1  < 1  < 1
Typical measurement time  s  < 1�5  < 1�5  < 1�5
External standard calibration value  g  2,000 (E2)  1,000 (E2)  1,000 (E2) 
(min� accuracy class)
Application range (temperature)    With “isoCAL” function: +5 ��� +40 °C  
    Without “isoCAL” function: +15 ��� +25 °C
Display result  
(depending on the set filter level)    0�1 – 0�4
Weighing pan size (W + D)   mm  140 + 140
Weighing chamber height (draft shield DE)
 mm  172
Protection     Protected against dust and water
* DIR  = Directive 2014/31/EU on non-automatic weighing instruments used within the European Economic Area 
**    =  According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing 
range� Typically the value can be higher depending on the setup location and the ambient conditions�
96  Cubis MSE Operating Instructions