Cubis MSE Operating Instructions  99
Technical Specifications
Model-specific Data
Verified models with EC Type Approval Certificate: Precision balances 
Modelle    MSE12201S-  MSE8201S- MSE5201S-  MSE70200S-  MSE36200S- 
   0CE  0CE  0CE  0CE  0CE
Accuracy class* K  K K  K  K
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d*  mg 100  100  100  1,000  1,000
Weighing capacity max*  g 12,200  8,200  5,200  70,200  36,200
Calibration value e*  g 1  1  1  10  1
Min� load min*  g 5  5  5  50  50
Tare equalization range (subtractive)  < 100% from max� weighing capacity
Application range according to DIR* g  5 – 12,200 5 – 8,200 5 – 5,200 50 – 70,200 50 – 36,200
Optimal starting point  
of the measuring range**  g 82  82  82  820  820
Typical stabilization time s < 0�8  < 0�8  < 0�8  < 1  < 1
Typical measurement time s < 1  < 1  < 1  < 1�2  < 1�2
Application range (temperature)    With “isoCAL” function: +5 to +40 °C | Without “isoCAL” function: +10 to +30 °C
Display result  
(depending on the set filter level)    0�1 – 0�4
Weighing pan size (W + D)   mm  206 + 206  206 + 206  400 + 300  400 + 300  400 + 300
IP protection   IP54
* DIR  = Directive 2014/31/EU on non-automatic weighing instruments used within the European Economic Area 
**    =  According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing 
range� Typically the value can be higher depending on the setup location and the ambient conditions�