Linked modules
1. Set up the system as illustrated in
Diagram 2. Link the required number of
modules by sliding the tongue and groove
edges together. Note the positioning of
the flow restrictor on the return line.
2. Use series interconnectors, (prod. no.
VFA031), to connect a single row of
modules. Detailed ordering information
for system components to be found of
page 7, Diagram 4.
3. In addition, use T-connectors, (prod. no.
VFA030), to connect two rows of modules
in parallel.
4. The flow path of the system can be con-
figured to suit the membrane|sample
combination. Most solutions are better
suited by modules connected in parallel.
Very low viscosity solutions favour mod-
ules connected only in series. Rinse the
system as detailed under single module
operation.