16
Recording and Evaluating Data
§ Press the key to clear the statistics memory prior to
initial data transfer
> Any previous values remaining will be evaluated first,
then you will obtain a printout of the statistics.
The statistics will be deleted.
§ Place the first sample on the weighing pan
§ To activate data transfer, press the key
> The readout shown on the left will appear
> “STAT” will be displayed for approx. 0.5 sec.
> This value will be read into the statistics memory as the
reference value
If you have deactivated the automatic taring function:
§ Remove the sample from the weighing pan
§ Place the next sample on the weighing pan
> The value will be automatically transferred when the
weighing instrument displays the stability symbol
> Each value will be printed out (example)
If you have activated the automatic taring function:
> The weighing instrument is automatically tared after initial
data transfer
§ Leave the sample on the weighing pan
§ Add the next sample to the one already on the weighing pan
§ Make sure that the expected total weight does not exceed
the maximum capacity of your weighing instrument
(otherwise, “H” will be displayed)
> The value will be automatically transferred when the
weighing instrument displays the stability symbol
> The value will be printed out
123456789012345678901234
001: + 123.6 g
002: + 117.8 g
003: + 119.1 g