15 Appendix - Allowable maximum speed
The maximum allowable speed needs to be reduced when centrifuging a solution with a density
greater than 1.2 g/ml. Failure to reduce the speed may result in damage to the rotor and
centrifuge. The revised maximum speed can be calculated with the following formula.
Example:
Where the density of the liquid is 1,9, the new allowable maximum speed would be calculated as
follows:
16 Appendix - RPM to RCF conversion
RPM can be converted in RCF with the following calculation.
RCF = 11.2 × r (RPM/1000)
2
or RCF = 1.12 × 10
-5
x r x (RPM)
2
Example:
Where the RPM is 12,000 RPM and radius of rotor is 6 cm, Rcf would be calculated as follows:
RCF = 1.12 × 10
-5
× (6) × (12,000)
2
= 9,651 x g