11 Troubleshooting
In the event of a malfunction, the error message with code number appears to indicate the
probable causes and the instrument will be in the stop mode. If other malfunctions without error
code indication occur, turn off the power immediately, identify the causes and follow the corrective
actions as recommended below. For any problems not covered here or you are unable to correct
the malfunction, contact LaboGene for assistance.
If the instrument stops due to an error indication, the run cannot be restarted until the error is
cleared. After the problem is corrected, reset the instrument to check if the error occurs again.
Error Indication
Possible Reason
ly
No display or power:
Power failure during operation;
display screen is blank
Corrective Action
1. Plug the power cord into the appropriate power outlet.
2. Turn the power switch off and back on to reset.

Corrective Action
1. Make sure to securely connect the power cord to the power outlet.
2. Turn the power switch off and back on to reset.

Corrective Action
Turn the power switch off and back on to reset.
* If problem cannot be resolved, use the Emergency Door Open function to
open the door manually to save samples. Call LaboGene for Service.

Operation cannot start
Corrective Action
Perform the recommended corrective action as listed in E1 and/or E9.

Corrective Action
1. Make sure to press down the door firmly until the latch handle is fully
retracted.
2. Turn the power off and back on to reset.

Corrective Action
Turn the power off and back on to reset.
* If problem cannot be resolved, use the Emergency Door Open function to
open the door manually to save samples. Call LaboGene for Service.

Corrective Action
Turn the power switch off and back on to reset.
* If problem cannot be resolved, call LaboGene for Service

Door does not open

Corrective Action
Turn the power off and back on to reset.
* If problem cannot be resolved, use the Emergency Door Open function to
open the door manually to save samples. Call LaboGene for Service.

Door does not close

Corrective Action
Turn the power off and back on to reset.
* If problem cannot be resolved, use the Emergency Door Open function to
open the door manually to save samples. Call LaboGene for Service.

Instrument is vibrating
excessively or making unusual
noise
Corrective Action
Perform the recommended corrective action as listed in E8.

Corrective Action
Turn the power off immediately and call LaboGene for Service.
* If problem cannot be resolved, call LaboGene for Service.