Turn the power switch off and back on to reset.
Failure to reach to 200 rpm
within 2 sec.
If problem cannot be resolved, it is possible that
Call LaboGene for Service
by the emergency door open function
Door opens during operation
Close the door and turn the power switch off and back on to
reset.
If problem cannot be resolved, it is possible that
Call LaboGene for Service
Detected internal temperature is
higher than 110
1. Clean the ventilation inlet opening or remove any objects are
blocking the vent.
2. Tun the power switch off and wait approximately 1 hour with
the door open to cool down the motor.
3. Turn the power switch back on to reset.
Turn the power switch off and back on to reset.
If problem cannot be resolved, it is possible that
amaged.
Call LaboGene for Service
Use Automatic Voltage Regulator to provide proper voltage.
tion allows
Use Automatic Voltage Regulator to provide proper voltage.