Turn the power switch off and back on to reset.
Detected actual speed value in
rpm is 1,000 rpm higher than set
speed value
If problem cannot be resolved, it is possible that
Call LaboGene for Service to update the software and tune revision
Turn the power switch off and back on to reset.
If problem cannot be resolved, it is possible that
Call LaboGene for Service to update the software and tune revision
1. Relocate instrument to a flat, level, and vibration free surface.
2. Turn the power switch off and back on to reset.
1. Make sure if samples are evenly weighted and distributed
symmetrically around the center of rotation.
2. Turn the power switch off and back on to reset.
1. Make sure the rotor and/or rotor lid is securely attached to
the shaft. 2. Turn the power
switch off and back on to reset.
Turn the power switch off and back on to reset.
Turn the power switch off and back on to reset.
If problem cannot be resolved, it is possible that
Call LaboGene for Service
1. Install the rotor as instructed in the manual and make sure
the rotor is aligned correctly.
2. Turn the power switch off and back on to reset.
1. Install the appropriate rotor.
2. Turn the power off and back on to reset.
Turn the power switch off and back on to reset.
If problem cannot be resolved, it is possible that
Call LaboGene for Service.
*The instrument MUST only be opened by a service engineer who is authorized by LaboGene.