Model Mini Centrifuge
04032013RS01
Cat. No. 9.900.900.730
CopyRight © 2013
Page 13 of 16
Error
Code
Problem
Possible Reason

Overspeed:
Corrective Action
Turn the power switch off and back on to reset.
E6
Detected actual speed value in
rpm is 1,000 rpm higher than set
speed value
If problem cannot be resolved, it is possible that

Call LaboGene for Service to update the software and tune revision
 error
E7
Control system failure
Corrective Action
Turn the power switch off and back on to reset.
If problem cannot be resolved, it is possible that

Call LaboGene for Service to update the software and tune revision

E8
Rotor Imbalance
Corrective Action
1. Relocate instrument to a flat, level, and vibration free surface.
2. Turn the power switch off and back on to reset.

Corrective Action
1. Make sure if samples are evenly weighted and distributed
symmetrically around the center of rotation.
2. Turn the power switch off and back on to reset.

Corrective Action
1. Make sure the rotor and/or rotor lid is securely attached to
the shaft. 2. Turn the power
switch off and back on to reset.

Corrective Action
Turn the power switch off and back on to reset.

Corrective Action
Turn the power switch off and back on to reset.
If problem cannot be resolved, it is possible that

Call LaboGene for Service
E9
Rotor recognition error

Corrective Action
1. Install the rotor as instructed in the manual and make sure
the rotor is aligned correctly.
2. Turn the power switch off and back on to reset.

Corrective Action
1. Install the appropriate rotor.
2. Turn the power off and back on to reset.
tem error
Corrective Action
Turn the power switch off and back on to reset.
If problem cannot be resolved, it is possible that

Call LaboGene for Service.
*The instrument MUST only be opened by a service engineer who is authorized by LaboGene.