Door lock failure (Emergency Door Open Function) 11.1
In case of power failure or any malfunction, the door will remain
locked as designed. Though, the door can be manually opened
with the Emergency Door Open Function.
Disconnect the power cord from the wall socket and allow the
instrument to come to a complete stop before opening.
Insert small screwdriver into the hole by sign on the
bottom plate of the instrument and move downwards to operate
the door latch.
mini, Microcentrifuge (220V, 50/60Hz)
includes Fixed Angle Rotor GRF-m2.0-12 & 12 adaptors for 0.2 mL
Fixed Angle Rotor, GRF-m2.0-12 with lid
PCR Rotor, GRA-s0.2-32 for PCR strips, 4-rows