Cat. No. 9.900.900.730
CopyRight © 2013
7. Carefully place the rotor onto the motor shaft. Hold the rotor with one hand and securely
tighten the rotor nut to the shaft by turning clockwise.
Turn the rotor nut counter clockwise to release the rotor from the shaft.
Check tubes and rotors for cracks and deformities before each use.
Do not attempt to open the lid unless the rotor has stopped completely.
Do not exceed safe rotor speed.
The operator should not leave the centrifuge until the full operating speed is
attained and the machine appears to be running safely without vibration.
Stop the centrifuge immediately and unplug the power cord if an unusual condition
(noise or vibration) begins.
If tube breakage occurs, turn centrifuge off immediately. Leave for 30 minutes to
reduce the risk of aerosols. The operator should wear proper gloves, remove
debris, clean and disinfect centrifuge interior and rotors.
Clean all spills immediately and decontaminate the instrument and rotor after use
with biological or radioactive materials.
1. can be found on the rear of the
instrument.
2.
3. Load sample tubes into the rotor symmetrically with evenly weighted samples and securely
close the rotor lid by pressing the snap connector on the rotor nut.
Do not attempt to exceed the maximum load or speed of the rotor. The maximum
allowable speed needs to be reduced to spin solutions with a density greater than 1.2
g/ml (see Appendix for details).
4. To set or change the centrifugation parameters, follow the instruction below.
a. Speed/Force: and press the arrow buttons
select or change speed in increments of 100 RPM from 1200-13500 RPM.
convert speed in RPM to RCF.
b. and press the arrow buttons
length of spin time from 1 ~ 30min. The remaining operation time will be displayed.
c.
hold the button for desired length of time and release to deactivate it.