5 Online sensor specifications
5.1 TS, FS and MZ series
TS series (models TS1, TSP, TSR, TSH, TSL, TSF, TSF1, TSW, TSB1, TSB2)
FS series (models FS1, FSP, FSR, FSH, FSL, FSW, FSB1)
MZ series (models MAZD, MBZD, MAZF, MBZF, MBZB, MZ1, MZH, MZB1)
Calibration: According to SCHMIDT factory procedure
Accuracy: For 10% to 100% of range:
TS: ± 1 % full scale
FS: ± 1.5 % full scale
MZ: ± 2 % full scale
MZ1, MZH, MZB1: 1.5 % full scale
Remainder of range and
Other calibration material: ± 3 % full scale or better
Overload protection: 100 % of range
Measuring principle: Strain gauge bridge
Measuring roller deflection: 0.5 mm max.
Signal processing: Analog
Output signal: 0 - 1 V DC (standard)
Damping (f
g
): Standard: 30 Hz (other values on request)
Temperature coefficient: Gain: less than ± 0.05% full scale/°C
Temperature range: 10 - 45 °C
Air humidity: 85 % RH, max.
Power supply: Standard: + 15 to + 24 V DC (21 mA, regulated)
Further technical details are provided in the instruction manual for the sensors.
5.2 SF series
Models SFZ and SFD
Accuracy: 0.5 % full scale
Max. applied force: 160 % full scale, overload protection afterwards
Overload protection: SFZ: 10 times of nominal load, max. 3200 N
SFD: 10 times of nominal load, max. 2000 N
Lateral force at the axis: Max. 100 % nominal load
Measuring principle: Strain gauge bridge
Output signal: SFZ: 5 - 20 N: 1 mV/V, 50 N and higer: 1.5 mV/V
SFD: 1 mV/V
Temperature range: 10 - 70 °C
Power supply: 10 V DC