EasyManua.ls Logo

SCHOTT TITRONIC universal - Performing Dosing; Incremental Dosing

SCHOTT TITRONIC universal
27 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38
Send results to
printer / computer (PC)
Fig. 22
short click
(0.3 seconds max.)
After pressing a key for the last time, wait at least
for 1 second, then press the grey key once
(3.0 seconds max.), and the displayed volume in
terms of ml will be sent to the printer / computer.
The display remains on the screen, titration can
be continued.
4.8 Dosing
Press the <MODE> key so many times until the
display shows “<MODE>:Dos”.
<Mode>:Dos
8.37
ml
Nom. vol. 15.00ml<F2>
Speed: 40.0ml/min<F3>
Dosing: new <Start>
The „Dosing“ function will eject at an exactly definable speed (rate) a nominal volume which is adjust-
able from 0.01 to 999.99 ml. You can set this dosing speed (rate) from 0.5 to 40.0 ml/min in the case of
the TITRONIC
â
ââ
â
universal / 20 ml Piston Burette, and to 0.5 to 99.9 ml/min in the case of the
TITRONIC
â
ââ
â
universal / 50 ml Piston Burette. For the 20 ml burette the max. rate is 40 ml/min, for the
50 ml burette it is max. 99.9 ml/min.
You set both the command volume and the dosing rate in a dialogue with the display of the
TITRONIC
â
ââ
â
universal Piston Burette, & „Setting the parameters“ chapter. The stored parameters
remain active even after the burette is switched off.
4.9 Incremental dosing
Use the <MODE> key to select the
„Incremental dosing“ function
You set the parameters in a dialogue with the bu-
rette, & „Setting the parameters“ chapter.
<Mode>:Increm.
0.00
ml
Increm.: 2.50 ml <F2>
End vol.: 50.00 ml <F3>
Delay : 10.0 s <F5>
Increment new: <Start>
With incremental dosing -Mode>:Increm- you divide an adjustable end volume -End vol.:- into individual,
adjustable volume steps -Increm:-, with a likewise adjustable waiting time -Delay- being inserted between
the individual steps. The addition of the individual volume steps is displayed as total volume.
Fig. 25 Minimal and maximal set parameters:
Example Parameter minimal maximal
Volume of the individual steps 2.5 ml Increment [ml] 0.01 99.99
Total volume 50 ml End volume [ml] 0.01 999.99
Waiting time after each step 10 s Waiting time [s] 0.5 999.9
= End volume
= 50 ml
Start
2.5 ml2.5 ml 2.5 ml
2.5 ml
10.0 s
Wait
10.0 s
Wait
10.0 s
Wait
If the increment is set to a value greater than the end volume, the burette will only dose 1 x “end volume”.
Fig. 23
Fig. 24

Related product manuals