41
Command Description Answer
99AAnn Transmit new address (nn = new address) nnY
aaABcommand Command after instruction AB ... is performed aaY
and sent to following device
aaBF Fill piston burette aaY
aaBN Reset ml display to zero aaY
aaBP Inquiry of piston position in scale parts aaBP=0 ... 8000
aaBV Inquiry of burette volume aaV=12.34
aaDA13.56 Dose with addition of volume aaY
aaDO123.4 Dose indicated quantity in ml aaY
aaDB123.4 Dose quantity in ml without filling and zero aaY
aaDH Endless dosing, to be stopped
aaDZ Start dosing, without filling; display will add up to max. volume 9999 ml aaY
aaGDM20,3 Dosing rate in ml/min aaY
aaGF30 Filling time in s aaY
aaKA Which key is being pressed? AaKey:“Text“
aaKD Keyboard disable aaY
aaKE Keyboard enable aaY
aaMO0 Inquiry of current function aaModeZ
Digit sequence: Z = 1 ð „Titration“
Z = 2 ð „Dosing“
Z = 3 ð „Incremental dosing“
aaMO1 Set titration function aaY
aaMO2 Set dosing function aaY
aaMO3 Set incremental dosing function aaY
aaRA Interchangeable unit - inquiry of volume size aaAUFSATZ:20
aaRH Identification request aaIdent:T universal
aaRC Send last command aa“last command“
aaRS Report status aaStatus:“text“
aaSC Continue after stop/continue function after stop aaY
aaSH Stop, burette stops aaY
aaSD1..8 Write string in n
th
matrix line: attach string to be written with “TEXT“ aaY
(limit characters to 21)
as a delimiter character to command e.g.: 01SD11”
(Please replace interchangeable unit)
aaST0 Delete display completely aaY
aaST1 Switch display fully on aaY
aaSR Stop current function aaY
aaSYS5 Set display language to German aaY
aaSYS1 Set display language to English -english- aaY
aaSYS2 Set display language to French -français- aaY
aaSYS3 Set display language to Spanish -español- aaY
aaVE Software version number aaVersion:99.23
aaVO Volume inquiry from display aaV=12.345ml