4.1 Location
1Place the centrifuge on a firm, flat and level table, ensure the four feet of this centrifuge stand on the
table firmly. Avoid installing on the slippery surface or surface prone to vibration.
2Ideal ambient temperature is 20℃±5, avoid placing the centrifuge in direct sunlight if temperature
exceeds 30.
3Keep clear of the centrifuge at least 10cm on both sides and at least 30cm behind it to guarantee the
cooling efficiency.
4Keep away from heat or water to avoid sample temperature issues or centrifuge failures.
4.2 Connection of the power cord and grounding
WARNING
To avoid electrical shocks, ensure your hands are dry when touching the power cord.
This centrifuge must be grounded properly.
A minimum 10A outlet providing a sufficient ground is required, and this must meet with local safety
requirements.