USER MANUAL
16
Two holes are located on the both sides in the top right corner of the cover.
Insert a screw driver into the hole and push forward to release the door.
10.3 Replacement of fuses
1There are two fuses, 250V, 6.3A time-delay type, size: Ф5×20.
2The fuse holder is located in the power inlet. Pull out the fuse holder from power inlet and replace
the fuses if necessary.
11. Instructions of rotor and tube
CAUTION
Read the instructions thoroughly, correct use rotor.
Do not exceed the allowable maximum speed of rotortube and adapters etc., be care that the allowable
maximum speed of some adapters are lower than the rotors maximum speed.
11.1 The rotor instructions
11.1.1 Rotor structure
Figure 11-1 The rotor structure
11.1.2 Available rotors and adapters
Closed angle rotors, such as AS30-2, AS18-5V, AS60-2, AS12-PCR8, AS6050V, are used for bio-safe
when the rotor lid was tightened with the rotor, centrifuge tubes will be enclosed into rotor to ensure the
sample does not leak in centrifugal process. If rotor lid is not available, the rotor would be no bio-sealing
function. The rotors can be used as follows:
Knob Locking nut Cover
Seal ring identify ring