USER MANUAL
Rotor type
ID
code
Tube/bottle
Adapter
Max.
speed
(rpm)
Max.centrifugal radius
r
max
(cm)
Max. RCF
Rcf (×g)
1
AS30-2
2/1.5ml ×30
6000
10
4020
0.2ml ×30
A02P2
6000
8.5
3415
0.5ml×30
A05P2
6000
9
3618
2
AS60-2
2/1.5ml ×60
6000
10
4020
0.2ml×60
A02P2
6000
8.5
3415
0.5ml×60
A05P2
6000
9
3618
3
AS18-5V
5mlV×18
6000
10
4020
4
AS12-PCR8
8-PCR ×12
6000
10
4020
0.2ml×96
6000
10
4020
5
AS6-50V
50mlV×6
6000
10.7
4306
6
A30-15
15mlV×30
4500
R1=14.2
R2=12.2
3210
2760
7
A8-50
15mlV×16
50mlV×8
5000
12.4
3460
8
SE4-100A
100ml ×4
4000
15.9
2840
85ml ×4
A85P100
4000
15.9
2840
50mlV ×4
A50VP100
4000
15.9
2840
15mlV ×8
A15VP100
4000
15.1
2700
310ml ×8
A10P100
4000
14.7
2630
9
S2-MP
(dimension)
mm
MTP
(128×85.6×15)
4000
12.1
2160
Cell culture
(128×85.6×21)
4000
12.1
2160
DWP
(128×85.6×45)
4000
12.1
2160
Kits
(128×85.6×60)
4000
12.1
2160
* 15mlV means 15ml conical tube, as well as 5mlV and 50mlV.
Table 11.1 Rotors and adapters
11.1.3 Notice
(1) The centrifuge rotor can separate sample which density lower than 2.0g/ml, if the samples density is
over 2.0g/ml, please calculate allowable speed depending on the following formula.
Allow Speed (rpm)= Maximum speed×(2.0(g/ml)/Sample density (g/ml))
1/2
2To prevent corrosion, remove the rotor from rotor chamber if do not use for a long time, then detach the
rotor lid and upside the rotor down to dry the tube holes.