USER MANUAL
18
(3)If some samples leaked in the rotor hole, wash the hole with water, apply a thin coat of silicon grease
on the rotor surface after drying.
(4)It is necessary for a regular maintenance for rotor, recommend to clean it each 3 months to keep
cleaning of tube hole and shaft hole, and then apply a thin coat of silicon grease on it.
11.1.4 Autoclaving
All rotors are manufactured in high-strength aluminum alloy material or stainless still and can be autoclaved:
121℃ (1.0kg/cm2), 20 minutes.
But some adapters are made of plastics, these adapters can be deformed after autoclaving, so you‟d better
use other disinfecting methods.
11.1.5 Bio-safe seal ring
The rotor is sealed by bio-safe structures, achieved using three high-temperature rubber seal rings. The seal
rings may fall off or aging after several autoclaving, need to be replaced or re-installed. The replacement
methods please refer to the section 9.3.
11.2 Tubes
11.2.1 Cleaning and sterilizing tubes
Condition Material
PA
PC
PP
Cleaning
Cleaning fluids
Acidic(pH5 or lower)
X
X
X
Acidic(higher than pH5 )
O
O
O
Alkaline(higher than pH9 )
O
X
O
Alkaline(pH9 or lower)
O
O
O
Neutral(pH7)
O
O
O
Warm water(up to 70℃)
O
O
O
Ultrasonic cleaning
Neutral detergent(pH7)
O
O
O
Sterilization
Autoclaving
115℃(0.7kg/cm
2
)30minutes
O
O
O
121℃(1.0kg/cm
2
)20 minutes
X
O
O
126℃(1.4kg/cm
2
)15 minutes
X
X
X
Boiling
15 to 30 minutes
O
O
O
Ultraviolet sterilization
200-300nm
X
X
X
Gas sterilization
Ethylene oxide
O
X
O
Formaldehyde
O
O
O