SD CodeFree™ BLOOD GLUCOSE MONITORING SYSTEM
64
Maintenance and Troubleshooting
accordance with local guidelines.
4) Control solution
If you are a medical professional or have the
control solution in your possession please note
the following.
• Keep the SD Control Solution in a cool dry
environment 8-30°C (46-86°F) .
• Do not refrigerate or freeze.
• Do not use SD Control Solution that has
passed its expiration date.
• TheSDControlSolutioncanbeusedfor3
months after opening the container. Do
not use after this time. Write the opened
date on the SD Control Solution container
when rst opened.
• No reconstitution or dilution is necessary.
• Wipe the container tip clean and reseal the
container tightly after each use.
4. Cleaning the Lancing Device
Clean the outside of the SD lancing device
regularly with 70% isopropyl (rubbing) alcohol.
Do not place the entire device under water. Do
not use bleach. At least once a week, disinfect
the removable comfort cap after cleaning by
placing it in 70% rubbing alcohol for 10 minutes.
Allow the cap to air-dry after disinfecting.