Maintenance seca
17.08.11/RLEU/ARI/MRE 13/36 30-34-00-812a
gain briefly press the tare/hold key (for
less than 1.5 seconds).
incremented by 1 g.
Repeat this step until the measured
weight matches the test weight (format:
xx.000 kg).
The display indicates the test weight
(format: xx.000 kg).
Press the tare/hold key until the display
shows “Sto”.
The display reads “Sto” and the scale
switches off automatically.
2.2 Adjusting the Head Slider [284|285]
The head slider electronics must be readjusted whenever the electronics are
replaced. Please refer to the instructions in the operating manual.
2.3 Displaying the Gravity Factor
[374|376|378|703|704|284|285|954|957|959|963]
To view the gravity factor, the scale must first be switched to adjustment mode
(see section 0 or 2.1.4).
As soon as the text “CAL” is displayed, press a key briefly (less than 1.5
seconds).
The display shows “GAL”.
Pressing a key for more than 1.5 seconds displays the gravity factor set. The
gravity factor is shown in mm/s².
The scale switches off again automatically after 18 seconds.
Note [376|378|704|285]:
On verified scales, the gravity factor can only be displayed as long as the calibration
count is 1.