Errors seca
17.08.11/RLEU/ARI/MRE 22/36 30-34-00-812a
Er:X:43 EEPROM defective Replace the
electronics
Er:X:45 EEPROM defective Replace the
electronics
Er:X:57 Quartz does not oscillate Replace the
electronics
3.6 Radio Error Message
[
374|376|378|703|704|284|285|954|957|959|963]
Display: Er:x:nr Electronics detecting a fault.
X : Module number
No Fault number
Scale displays Cause Remedy
Er:X:71
Radio module is
switched off
Switch on the radio
module via the
menu (see
operating
instructions)
Er:X:72
Error in teach-in mode Repeat teach-in for
the devices (see
operating
instructions)