Errors seca
17.08.11/RLEU/ARI/MRE 21/36 30-34-00-812a
Er:X:32 Command buffer
overflow error
Check the bus
connection/
modular cable (if
provided)
Isolate briefly (60
sec)
Er:X:40 EEPROM data faulty
Readjust the scale
or upload the scale
adjustment data to
the EEPROM via
seca serva 2.0
Er:X:41 Incorrect EEPROM
access
Replace the
electronics
Er:X:43 EEPROM defective Replace the
electronics
Er:X:44 Calibration counter faulty Replace the
electronics
Er:X:45 EEPROM defective Replace the
electronics
Er:X:57 Quartz does not oscillate Replace the
electronics
3.5 Head Slider Error Messages [264|274|284|285]
Display: Er:x:nr Electronics detecting a fault.
X : Module number
No Fault number
Scale displays Cause Remedy
Er:X:40 EEPROM data faulty Replace the
electronics
Er:X:41 Incorrect EEPROM
access
Replace the
electronics