Maintenance seca
17.08.11/RLEU/ARI/MRE 15/36 30-34-00-812a
2.4.1 Example for Gravity Factor Setting [374|703|284|954]
The current gravity factor shown on the display is 98135 kg mm/s².
The target gravity factor is 98150 mm/s².
In this example, you must switch the scale to incrementing mode to set the
gravity factor. When the tare/hold key has been pressed (for more than 1.5
seconds), "inc" starts flashing.
Now start to increase the value by 5 mm/s² (by briefly pressing the tare/hold
key). After this has been repeated three times, the value 98150 mm/s² appears
on the display.
Complete the gravity factor setting procedure as described above.
2.4.2 Summary of a Typical Gravity Factor Setting Sequence
[374|703]
Action carried out by the user
Result
The scale is switched off. No values or characters on the display.
Press the tare/hold key and keep it
pressed while simultaneously pressing
the start key.
The contents of the calibration counter
flash on the display for 18 seconds.
Release both keys. The contents of the calibration counter
flash on the display for 18 seconds.
Press the tare/hold key for more than 1.5
seconds.
The system switches to adjustment
selection mode and “CAL” flashes on the
display.
18 seconds have passed and no key has
been pressed.
The scale switches off.
Briefly press the tare/hold key (for less
than 1.5 seconds).
“GAL” flashes on the display.
Press the tare/hold key for more than 1.5 "dec" appears on the display (flashing)
and then the current gravity factor - also