Maintenance seca
17.08.11/RLEU/ARI/MRE 16/36 30-34-00-812a
seconds. flashing.
Decide whether to decrement or to
increment:
Incrementing:
Press the tare/hold key for more than 1.5
seconds.
Decrementing: No action required as the
scale already is in dec mode.
Incrementing mode is activated; "inc"
appears on the display.
No output on the display.
Decrementing or incrementing:
A
gain briefly press the tare/hold key (for
less than 1.5 seconds).
The gravity factor displayed is
decremented or incremented by 5 mm/s².
Repeat this step until the gravity factor
displayed matches the target gravity
factor.
The display shows the gravity factor
(format: xxxxx mm/s²).
Press the tare/hold key until the scale
switches off.
The display reads “Sto” and the scale
switches off automatically.