EasyManua.ls Logo

Seca 374 Series - Page 25

Seca 374 Series
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Measurements seca
17.08.11/RLEU/ARI/MRE 24/36 30-34-00-812a
Messung zwischen Ohmscher Widerstand []
V
+
und Sig
+
4 (grün und rot) ~ 800-850
V- und Sig
+
(grün und weis) ~ 800-850
V
+
und Sig- (rot und schwarz) ~ 800-850
V
-
und Sig- (weiß und schwarz) ~ 800-850
Sig
+
/Sig-/V
+
/V- und Alukörper
der Wägezelle
> 500 M
Table 2: Resistance of Flintec / Kogean Planar Beam / --Wing load cell [376|378|284|285]
Measurement between Ohmic resistance []
V
+
and Sig
+
4 (green and red) ~840
V- and Sig
+
(green and white) ~840
V
+
and Sig-
(red and black) ~840
V
-
and Sig- (white and black)
~840
Sig
+
/Sig-/V
+
/V- and aluminium
body of the load cell
> 500 M
Table 3: Resistance of VCARE [374]
The relevant cabling diagram shows the designations and associated cable
colour. The measured values only have a limited informative value, however, as
even a defective load cell can supply absolutely correct values when no load is
placed on it. If the load cell supplies incorrect values when unloaded, it is
definitely defective.

Table of Contents

Related product manuals