8. WHAT TO DO IF …?
Malfunction Cause/Remedy
… no weight is displayed
during weighing?
The scales are not connected to the power
supply
- Check the scales are switched on
- Check batteries are inserted
0.00 does not appear
before weighing?
A load was placed on the scales before being
switched on.
- Remove load
- Switch scales off and back on again
… a segment lights up
continuously or not at
all?
There is a fault at that point.
- Notify maintenance service.
is displayed?
Battery voltage is falling off.
- Insert new batteries as soon as possible
bAtt is displayed?
Batteries are flat.
- Insert new batteries
St0p is displayed?
The maximum load has been exceeded.
- Reduce load
is displayed?
The ambient temperature of the scales is too high
or too low.
- Set up scales in an ambient temperature
between +10 °C and +40 °C
- Wait approx. 15 minutes until scales have
adjusted to ambient temperature
... if after switch-on,
measured results are
transmitted for the first
time and two acoustic
signals are heard?
The device was unable to send measured
results to the radio receiver (seca wireless
printer or PC with seca USB wireless module).
Check that the scales are integrated in
the wireless network.
Check that the receiver is switched on.
Nearby RF equipment (e.g. mobile phones) are
interfering with reception.
- Make sure that RF equipment is kept at least
1 metre away from transmitters and receivers
in the seca wireless network.
NOTE:
If such interference is not eliminated, no
further acoustic warning will be given with
subsequent attempts at transmission.